model | RWS-G01 | Equipment parameters | Device configuration |
Power Supply | Voltage | 380±10V/50H Z ; | 16A/3 Hole plug ten Meter power cord |
Rated power | six kW |
Microwave system | Microwave output power | zero ~3 kW | one point five KW microwave system 1 set, including water cooling Magnetron And all copper leakage transformer |
Microwave output frequency | 2.45GHz±25MHz |
Thermal insulation system | Rated temperature | one thousand two hundred and fifty ℃ | contain one thousand and six hundred 1 set of thermal insulation structural formwork at ℃/1800 ℃, 2 quartz tubes, 4 quartz boats, 8 quartz crucibles |
Continuous operating temperature | one thousand one hundred and fifty ℃ |
Heating space | φ forty × 130mm dia × Width) |
heating rate | Rapid heating, average heating rate ≥ 40 ℃/min |
Control system | Temperature measurement mode | Infrared temperature instrument | Including thermocouple one Branch, infrared thermometer one set |
Measuring range | zero ~1300℃ |
control accuracy | < ±3℃ |
Control system | Automatic control, manual control and constant temperature control | forty The section can be set with process parameters, data storage and experimental data export; Manual, automatic, constant temperature control mode, real-time curve display; Dynamic data screensaver | contain PLC1 Pieces seven Inch touch screen one Component, contactor one Set, circuit breaker one Set, relay one Sleeve and voltage regulating module one Set, switching power supply one Sleeve, signal isolator one Set USP Interface one individual |
Atmosphere system | Static vacuum degree | <50Pa | Including vacuum pump one Table, vacuum gauge one Support and pressure gauge one Support and vacuum flange two Set, atmosphere control pipeline two Sleeve, vacuum valve three set |
Climate requirements | Fast pre vacuum, oxidizable gas, inert gas, weak reducing gas, etc |
safety system | Prevent microwave leakage | Microwave leakage level<0.1Mw/ cm two | Stainless steel furnace door, microwave fully shielded spring ring one Set and furnace door closing protection switch one Handheld microwave leak detector one individual |
Magnetron alarm | Over temperature alarm, over-current alarm |
Furnace door protection alarm | Furnace door closing protection switch |
Circulating cooling water | mode | External circulating water or tap water | Flowmeter one Set Water load one set |
pressure | ≥0.15MPa |
flow | ≥1.5m three /h |
Inlet water temperature | 5-30℃ |
Outlet water temperature | ≤60℃ |
Total hardness of circulating cooling water | <60mg/L |
Overall dimensions of equipment | Overall dimensions | one hundred and fifty × eighty × 176 cm (long × wide × High) | Carbon steel plastic spraying cover, mobile carbon steel plastic spraying test bench one piece |
Furnace tube | φ seventeen × 68㎝ |
area covered | ≤0.5㎡ |